Laparoscopy and Ventilators Suppliers in Salem

FRESENIUS DIALYSIS MACHIE-2008K(LCD TOUCH SCREEN)

Specifications :


  • Power Consumption : Does not exceed 12.5amps
  • Water Supply : Min 20psi and max 105psi
  • Water Quality : AAMI Standard
  • Drain : 3 Feet Maximum height
  • Dialysate Flow : 300/400/500/600/700/800ml/min
  • Proportional Mixing : Selectable 1:34/1:44/1:35:83 Volumetric,130 to 155 meq/L Na
  • Bicorbonate Setting : Selectable 1:27.6/1:19.1?
  • Dialysate Heating : 34 to 39"C
  • Heat Disinfection : 83:c
  • Blood Pump : 0 - 600ML/MIN
  • Heaparin Pump : 30ML Disposable Syringe
  • Venous Pressure Monitor : -80 to 500mmhg with automatic set window
  • TMP : +60 to -500mmhg
  • Level Detector : Ultrosonic Impuse Sensor
  • Blood Leak Detector : Two color light source transmitter
  • Ultra filtration control : Volumeteric Control 0 - 4000ml/min
  • UF Time : Digital Display
  • UF Goal : Digital Display
  • Additional Monitoring : Alarm in case of power failure